Specifications
|
Current Measurement
|
Range:
|
0 to 24 mA
|
|
Resolution:
|
0.001 mA
|
|
Accuracy:
|
0.025% Rdg + 2 LSD
|
|
Current Sourcing
|
Range:
|
0 to 20 mA or 4 to 20 mA
|
|
Accuracy:
|
0.025% Rdg + 2 LSD
|
|
Drive Capability:
|
1200 Ω; 950 Ω at 20 mA in HART mode
|
|
Current Simulation
|
Range:
|
0 to 20 mA or 4 to 20 mA
|
|
Accuracy:
|
0.025% Rdg + 2 LSD
|
|
Ext. loop voltage requirement:
|
12-30 V
|
|
Loop Power
|
While measuring mA, 24 V
|
|
Voltage measurement
|
Range:
|
0 to 28 V DC
|
|
Resolution:
|
1 mV
|
|
Accuracy:
|
0.025% Rdg + 2 LSD
|
Environmental Specifications
Operating Temperature -10ºC to 55ºC
Storage Temperature -30ºC to 60ºC
Humidity (Without Condensation) 0 to 90% (0ºC to 35ºC); 0 to 70% (35ºC to 55ºC)
Temperature Coefficient -10ºC to 18ºC, 28ºC to 55ºC, ±.005% of range per ºC
Safety Specifications
Maximum voltage 30V
Mechanical & General Specifications
Size 152 x 70 x 41 mm
Weight 224 g (340 g including holster)
Batteries One 9V battery Alkaline(YHS-707) , One 9V battery
Warranty Two years
Shock & Vibration per MIL-T-28800 for a Class 2 Instrument
Display current and % of span Yes
Auto step, auto ramp Yes
Span Check Yes
| Payment Terms: | T/T,Moneygram,Westunion |
click for larger image
Img 1
This member assumes full responsibility for the content of this listing. DIYTrade accepts no responsibility whatsoever in respect of such content.
To report fraudulent or illegal content, please
click here.