Model: | HX-801E |
---|---|
Brand: | Huaxi |
Origin: | Made In China |
Category: | Electronics & Electricity / Other Electrical & Electronic |
Label: | Infusion pump , Volumetric pump , Drip rate pump |
Price: |
US $600
/ pc
|
Min. Order: | 1 pc |
Live Chat: | Last Online:10 Nov, 2010 |
1. IV Set Management and Flow Rate Calibration
IV sets can be added, stored and deleted in memory, and easily used after calibration.
2. Optional Infuse Rate Unit
mL/h and volume/time
3. Infusion Mode
Volumetric
4. Pumping Mechanism
Peristaltic finger pump
5. IV Sets
Standard IV sets of 10, 15, 20 and 60d/mL, with diameter 3.4 ~
6.
1.0 ~ 1200 mL/h
7. Increment
From 0.1 mL/h to 99.9 mL/h, 0.1 mL increment; ≥100 mL/h, 1 mL increment.
8. Flow Rate Accuracy
≤ ± 5% (after calibration)
9. Volume Limit
1 ~ 9999 mL
10. Volume Infused Display
0 ~ 9999 mL & >9999 mL
11. Purge Rate
600 mL/h
12. KVO Rate
1 mL/h
13. Bolus Mode
Available for Bolus mode, flow rate adjustable.
14. Occlusion Pressure
Pressure detection device with 8 sensitivities setting, 40 ~ 140KPa.
15. Air-in-line Detection
Air bubble detected by ultrasonic and the device with 8 sensitivities setting, 50 ~ 1000μL.
16. Alarm Volume
8 levels setting
17. Anti-reverse Function
Anti-reverse function on motor to prevent upstream
18. Self-check
Self-check when start up and during infusion
19. Audio and Visual Alarms
Air-in-line, Door open, Occlusion, Infusion finished, Pause overtime, Leakage, Empty, Low battery,
20. Display
Blue LCD with high brightness
21. Power Source
AC: 100V ~ 240V, 50Hz/60Hz, ≤ 25VA
22.
Rechargeable Ni-MH battery: DC12V, 2000mAh, life: ≥ 4 hours at 25mL/h
23. Classification
Class Ⅱ, CF, IPX4
24. Dimensions and N.W.
Payment Terms: | T/T |
---|---|