ConST316 Multifunctional Temperature Calibrator
An innovative multifunctional calibrator, ConST316 contains powerful task management function. Through smart phone-type operation, ConST316 is definitely your good helper for your effective and convenient calibration.
Outstanding Functions
u Smart phone-type operation, icon-based menu management
u Strong tasks management functions: supporting Information management of gauges under calibrate、Parameter setup while calibration、Auto-executed while calibration、Data automated analysis、AutoMark of overproof point
u Fast saving calibration results、and Tasks download、upload available;
u Thermal units exchange tool, which conforms to ITS-90;
u Advanced automated cold side compensation technique: Build-in cold side heat preservation block; temperature variation tracing rapidly; and take the lead in realizing the calibration of cool side sensor.
u Reliable mis-operation protection technique: 30V mis-operation voltage can be bore between any two jacks. The test jack can bore 1A mis-operation current for quite a long time. After the mis-operation stopped, the calibrator will return to normal and no need to change fuse.
Other Functions
u screen snapshot file saving is available.
u Supports the calibration of square root transmitter and its display instrument;
u Pulse frequency output is available;The calibration of Flow Totalizer is convenient.
u Used as high accuracy platinum resistance digital thermometer. Supports parameter adjustment of Ro, a, b and c.
u The rechargeable battery can be replaced and quite a long continuing working time is available.
u Introducing 3.5 inch TFT color screen, English display.
Specification
Measurement(working environment: 20℃±5℃)
|
Type
|
Range
|
Resolution
|
Accuracy
|
|
MV Voltage
|
(-75.0000~75.0000)mV
|
0.1µV
|
±(0.01%RD+0.005%FS)
|
|
Voltage
|
(-30.0000~30.0000)V
|
0.1mV
|
±(0.01%RD+0.005%FS)
|
|
Current
|
(-30.0000~30.0000)mA
|
0.1µA
|
±(0.01%RD+0.005%FS)
|
|
Frequency
|
(1~50000.0)Hz
|
0.1Hz
|
±(0.005%RD+0.002%FS)
|
|
Pulse
|
0~999999
|
1
|
±1
|
|
Resistance
|
2W
|
(0~400.000)Ω
|
1m Ω
|
±(0.02%RD+0.005%FS)
|
|
3W
|
(0~400.000)Ω
|
1m Ω
|
±(0.02%RD+0.005%FS)
|
|
4W
|
(0~400.000)Ω
|
1m Ω
|
±(0.01%RD+0.005%FS)
|
|
Pressure
|
Range and Accuracy refer to CDP intelligent pressure module selection
|
Output (working environment 20℃±5℃)
|
Type
|
Range
|
Resolution
|
Accuracy
|
|
MV Voltage
|
(-10.000~75.000)mV
|
1µV
|
±(0.02%RD+0.005%FS)
|
|
Voltage
|
(0.0000~12.0000)V
|
0.1mV
|
±(0.02%RD+0.005%FS)
|
|
Current
|
(0~22.000)mA
|
1µA
|
±(0.02%RD+0.005%FS)
|
|
Frequency
|
(0~50000.0)Hz
|
0.1Hz
|
±(0.005%RD+0.002%FS)
|
|
Pulse
|
0~999999
|
1
|
±1
|
|
Resistance
|
(1~400.00)Ω
|
10mΩ
|
±(0.02%RD+0.005%FS)
|
|
DC24V
|
|
|
24V±0.5V
|
General Specification
|
Connection Port
|
RS232
|
|
Working Environment
|
(-10~50)℃
|
|
Saving Environment
|
(-20~60)℃
|
|
Relative Humidity
|
<90%RH,non-gel
|
|
Pressure
|
(86~106)kPa
|
|
Power Supply
|
Rechargeable battery or AC adaptor
|
|
Battery
|
Polymer Li-ion battery,charging time less than 4 hours,continuing working: more than 15 hours(24V power supply/12mA)
|
|
Size
|
192mm × 100mm × 54mm
|
|
Weight
|
720g
|
Simulating thermal couples Accuracy(working environment: 20℃±5℃)
|
Thermocouple Type
|
Temperature range (℃)
|
Specified Temperature (℃)
|
Related Accuracy(℃) Measure Source
|
|
S
|
-50 ~1768.1
|
-20~100
|
1.6
|
1.2
|
|
100~400
|
1.2
|
0.8
|
|
400~1768.1
|
1
|
0.6
|
|
R
|
-50 ~1768.1
|
-20~100
|
1.6
|
1.2
|
|
100~500
|
1
|
0.8
|
|
500~1768.1
|
0.8
|
0.6
|
|
B
|
0 ~1820
|
600~900
|
1.3
|
1
|
|
900~1200
|
1
|
0.8
|
|
1200~1820
|
0.8
|
0.6
|
|
K
|
-270 ~1372
|
-200 ~ -100
|
0.6
|
0.4
|
|
-100~600
|
0.3
|
0.2
|
|
600~1372
|
0.4
|
0.4
|
|
N
|
-270 ~1300
|
-200~-100
|
0.8
|
0.6
|
|
-100~1300
|
0.4
|
0.3
|
|
E
|
-270 ~1000
|
-200~-100
|
0.4
|
0.3
|
|
-100~0
|
0.2
|
0.2
|
|
0~700
|
0.2
|
0.2
|
|
700~1000
|
0.2
|
0.2
|
|
J
|
-270 ~1200
|
-210~ -100
|
0.5
|
0.3
|
|
-100~1200
|
0.3
|
0.3
|
|
T
|
-270 ~400
|
-200~0
|
0.6
|
0.4
|
|
0~300
|
0.2
|
0.2
|
|
300~400
|
0.2
|
0.2
|
Simulating thermal resistance accuracy(working environment: 20℃±5℃)
|
Thermal resistance
|
Temperature range (℃)
|
Specified Temperature
|
Accuracy (℃)
|
|
Measure 2W/3W 4W
|
Source
|
|
Pt100
|
-50 ~1768.1
|
-100~200
|
0.14
|
0.09
|
0.09
|
|
200~600
|
0.25
|
0.15
|
0.15
|
|
600~1850
|
0.34
|
0.21
|
0.21
|
|
Pt10
|
-50 ~1768.1
|
-100~200
|
0.68
|
0.64
|
0.64
|
|
200~600
|
1.18
|
1.09
|
1.09
|
|
600~850
|
1.71
|
1.57
|
1.57
|
|
Cu100
|
-50 ~150
|
-50~150
|
0.44
|
0.4
|
0.4
|
|
Cu50
|
-50 ~150
|
-50~150
|
0.51
|
0.47
|
0.47
|
click for larger image
Img 1
Img 2
This member assumes full responsibility for the content of this listing. DIYTrade accepts no responsibility whatsoever in respect of such content.
To report fraudulent or illegal content, please
click here.
China Suppliers Quick Searching:
,
,